Q Exactive Plus (Q)


The Thermo Scientific Q Exactive Plus benchtop LC-MS/MS combines high-performance quadrupole precursor selection with high resolution, accurate-mass (HR/AM) Orbitrap detection to deliver high performance and tremendous versatility. With a fast scan speed and multiplexing capabilities, the Q Exactive Plus mass spectrometer is an outstanding detector for fast chromatography separation techniques. The superior quality of Q Exactive Plus MS/MS data enables identification and quantitation of more compounds with greater confidence. The Q Exactive Plus LC-MS/MS system not only offers broad screening capabilities but also excels at targeted quantitation experiments.

Visit the Planet Orbitrap website for more information

11/19/2013 QE Plus (SN03039L / UW tag 2019655) Installed
6/6/2012 QE (SN01284L) Installed, 11/14/2013 retired

Specifications

Back to Top

    Scan Rate
  • Up to 12 Hz at resolution setting of 17,500 at m/z 200
  • Multiplexity: up to 10 precursors/scan
    Resolution
  • 280 000 at m/z 200 option on the QE Plus
  • 140 000 at m/z 200 at a scan repetition rate of 1 Hz
    Mass accuracy
  • < 3 ppm RMS error with external calibration
  • < 1 ppm RMS error with internal calibration
    Sensitivity
  • Intra-scan dynamic range > 5000:1
  • Full MS: 500fg Buspirone on column S/N 100:1
  • SIM: 50fg Buspirone on column S/N 100:1
    Features
  • Higher Energy Collisional Dissociation (HCD)
  • S-Lens ion source for increased sensitivity
  • Quadrupole mass filter
  • Orbitrap mass analyzer
  • Spectral multiplexing for enhanced duty cycle

Instrument control software

Back to Top

Instrument calibrations

Back to Top

Method parameters

Back to Top

Method parameters for peptide identificaion on the QEplus
Check out this Technical Guide on Thermo's Planet Orbitrap website
How to best utilize your QE/QE Plus for maximum peptide IDs and for peptide quantitation for more information

For bottom-up proteomics method (DDA) use the Full MS followed by ddMS2 mode
Method ParameterRecommended Setting
Chrom. peak width dictates how often the AGC pre - scan is collected, adjust this time based on your chromatographic peaks widths. e.g. for narrow peak widths it is a critical for the instrument to take more frequent pre - scans based on the dramatic changes in ion flux
MS1 resolution for regular DDA 70k is recommended; for very complex samples higher resolution might be beneficial to resolve close m/z ions
MS2 resolution 17.5K is the lowest resolution setting, therefore the fastest scan rate. For tryptic digests, this resolution is usually adequate for MS2 scans since the spectra are less complex and fragment ions are mostly +1 ions
MS1 AGC target 1e6 with a max IT of 30ms is recommended
MS2 AGC target 5e4 with a max IT of 50ms is recommended to maximize scan speed
1e5 with a max IT of 100ms if you load <100ng
Loop count Refers to how many MS2 scans will be acquired between each full MS scan;
we use 20, Thermo recommends 10 for 20-30 sec wide peaks
Isolation window 2 Da on the QE
1.5 Da on the QEplus, the new segmented quad improves isolation and lowering the isolation window to 1.5 Da does not affect sensitivity, but can reduce the amount of interferences and increase the number of identificaions
NCE Normalized collision energy, 25 (28 for iTRAQ samples)
applied voltage (eV) depends on the NCE setting, the m/z and the charge state factor.
If you did want to know what the exact eV being applied you can view that in the scan header of each MS2 scan in qual browser.
Intensity threshold 1.0e4 (use the underfill % to set the int. threshold) minimum precursor intensity to trigger a MS2 scan
the MS2 target value, maximum injection time and underfill ratio ALL determine the threshold
Peptide Match we usually leave it off, but Thermo recommends to use preferred
Equivalent setting as "monoisotopic precursor selection" MIPS, to select ions with peptide-like isotopic distributions
On: will only select ions with peptide-like isotopic distribution
Preferred: ions with peptide-like isotopic distribution are triggered with preference, but will select other ions for dd-MS2 after all ions with "peptide match" were selected
Exclude Isotopes On is recommended
this parameter is de-coupled from "peptide match" or MIPS. So regardless of what your setting is for peptide match, you can still exclude isotopes
Dynamic Exclusion 30 sec, In complex mixtures, increase dynamic exclusion to reduce repeat sampling and maximize number of unique peptides
Default Exclusion width is ∓ 10ppm, but the exclusion widths for dynamic excl, inclusion list, exclusion list, etc are all changeable parameters

Method parameters for peptide quantification on the QE plus
The sacrifice in targeted experiments: If you want to quantitate as many targets as you can, well then you will be sacrificing depth and most likely overall sensitivity.
On the flip side, if ultimate sensitivity is your priority, then you have to sacrifice the number of targets you want to quantify.

Check out our PRM overview page for more information

TMT instrument QE plus settings
Back to Top

The table below lists instrument method settings recomended by Thermo for the Lumos with Tune 3.3.
with APD (Advanced Peak Detection) on
based on this document: TMT/TMTpro Instrument Acquisition Parameter Settings

Properties QE classic 120 min TMT 11plex QE classic 120 min TMT 16plex QE plus 120 min TMT 11plex QE plus 120 min TMT 16plex
Resolution Full MS 70,000 70,000 70,000 70,000
AGC target Full MS 3e6 3e6 3e6 13e6
MS max IT, ms 50 50 50 50
Scan range, m/z 350 - 1500 350 - 1500 350 - 1500 350 - 1500
Loop count 15 10 15 10
MS2 max IT, ms 250 250 120 120
MS2 Isolation window, Th 1.2 1.2 0.7 0.7
MS2 NCE, % 32-34 28-32 32-34 28-32
MS2 Intensity threashold 1e4 2e4 1e4 2e4
Dynamic exclusion, s 45 45 30 30
MS2 Resolution 35,000 35,000 35,000 35,000
MS2 AGC target 1e5 2e5 1e5 2e5
Peptide match preferred preferred preferred preferred
First mmass, m/z 110 110 110 110

Resolving Power and Transient Length

Back to Top

Here is a full table of the available resolution settings. Note: higher resolution does not always result in better mass accuracy. The higher the resolution increases your ability to distinguish between m/z ions. As you can see in this table, there is always a trade of higher resolution with time. 17.5 is the lowest resolution setting, with the fastest scan speed of ~13 Hz. On the QE, the highest resolution setting is 140K that is 1.5 Hz.
On the QE plus with enhanced resolution mode, you can go up to 280K, which is a less than 1 Hz.
To fully take advantage of the parallel fill and detect capabilities of the Q-Exactive, we need to balance the max fill times with the transient length. For resolving power 17,500, detection time is about 64 ms. There is always a small amount of inter-scan delay. Up to 50 ms can be used to fill the C-trap with ions- and without increasing cycle time.

Res. at m/z 200Res. at m/z 400 Transient length [ms]Approx. scan speed [Hz]"Free" fill time [ms]
17,500 12,500 64 13 50
35,000 25,000 128 7 110
70,000 50,000 256 3 240
140,000 100,000 512 1.5 500
280,000 200,000 1024 < 1 1000

HPLC setup

Back to Top


the following documents are guidelines to setup the capillary trap/column on our LC-MS systems, this is intended as a supplement to the hands on training required to use the UWPR instruments.

Instrument Settings

Back to Top

The table below lists some of the instrument method settings we tested on our QEplus.
We use a nanoAcquity UPLC with home made columns (30cm x 75μm, Magic C18AQ 100Å 5μm) and trap column (3cm x 100μm, Magic C18AQ 200Å 5μm), with 5-30% B (ACN, 0.1%FA) in 90min gradient. The sample is a Ramos cell whole cell tryptic digest (Thanks to Dr. Richard G. James, Seattle Children's), 100 ng on column injection. We typically get ~19000 unique peptides with Comet search and peptide prophet p>0.9,
The results below are based on the number of unique peptides with Comet search and peptide prophet p>0.9.

Q Exactive plus
Best Result (100%)Tested Values (X%)
Software Version QE 2.3
Include: 2-5 charge (MS1 35k) Include: unassigned & 2-5(89%), 2-5 (100%), unassigned (71%)
Include: 2-5 charge (MS1 35k) Only 2 plus (65%), 3 plus(43%), 4 plus (7%)
MS1 resolution 70k MS1 resolution 35k (97%), 140k (98%)
Isolation width 1.6 Isolation width 2 (97%)
Complex mix DDA Blank following complex mix (0.6%)
DDA: Top 30 DDA: Top 10 (99%), Top 20 (98%), Top 30 (100%), Top 40 (97%)
DDA: Top 30 4GPF windows (80%), MS1 scan split in 4 mz windows (84%); windows for both were 400-610, 600-810, 800-1000, 1010-2000
Software Version QE 2.5
NCE 25 NCE 25 (100%), 28(94.9%), 30 (90.0%), 32 (78.7%)
MS1 m/z 400-2000 400-2000 (100%), 400-1500 (97.1%)
Iso Width 1.6 Iso Width 0.7 (80.2%), 1.2 (94.8%), 1.6 (100%)
Fusion vs Lumos vs QEplus (OTMS2) Fusion (SW1.2) 15900 (89.8%), Lumos (SW2.0 beta) 17700 (100%), QE (SW2.5) 15760 (89.0%)
Software Version QE 2.8 SP1
minAGC (50ms IT) 5e2 100% 1e2 (99.5%), 2.5e2 (99.2%), 5e2 (100%), 1e3 (99.6%), 5e3 (98.1%), 1e4 (97.1%), 5e4 (74.8%)
= IntThreshold 1e4 (100%) 2e3 (99.5%), 5e3 (99.2%), 1e4 (100%), 2e4 (99.6%), 1e5 (98.1%), 2e5 (97.1%), 1e6 (74.8%)


The table below lists some of the instrument method settings we tested on our QE (before the QEplus upgrade)
Same setup except the sample is a yeast whole cell tryptic digest (Thanks to Dr. Judit Villen, UW), 100 ng on column injection. We typically get ~10000 unique peptides with sequest search and peptide prophet p>0.9,
The results below are based on the number of unique peptides with Sequest search and peptide prophet p>0.9.

Q Exactive
Best Result (100%)Tested Values (X%)
Isolation width 2 Isolation width 1 (95%)
Excl.IsoOn All off (99%), Apex trigger (91%), PepMatchOn (88%), PepMatchPref (62%)
Only Excl.IsoOn Exl.IsoOn and PepMatchOn (95%), Exl.IsoOn and PepMatchPref (97%)
DDA top 20 DDA top 30 (97%)
Signal Threshold 10k 2k (96%), 5k (94%), 10k (100%), 20k (93%)
MS2 target value 5e4 5e4 (99%), 1e5 (98%), 2e5 (98%), 5e5 (97%), 1e6 (100%)
MS1 target value 1e6 1e5 (90%), 5e5 (94%), 1e6 (100%)
iTRAQ NCE peptide ID's:NCE 20 (80%), NCE 25 (100%), NCE 30 (90%), SNCE 20/25/30 (83%)
Missing ITRAQ reporter (% of total ID's) all channels   113     114     115     116     117     118     119     121
NCE 20     2.8               10.4     10.2     11.6     16     11.7     9.1     20.6     7.5
NCE 25     0.3                 1.3       1.1       1.5       2.4     1.6     1.1     3.5       0.7
NCE 30     0.0                 0.1       0.2       0.2     0.4     0.3     0.1     0.8     0.0
SNCE 20/25/30     0.5                 2.5       2.3       2.8     4.2     2.8     1.9     5.7       1.7
Loading amount on column 1000 ng (96.2%), 500 ng (100%), 250 ng (96.7%), 100 ng (96.8%)

Hela digest concentration curve

Back to Top

Different concentrations of Pierce HeLa Protein Digest Standard (50, 100, 200, 500, 800, 1000 ng) were injected and separated with a 90 min 5-35% acetonitrile gradient over 90 min. Peptides were identified with our standard DDA method and search tools.


TIC (total ion chromatogram) of the different concentrations overlay, normalized to 1000 ng chromatogram.



    Useful resources

    Back to Top

  • Q Exactive Tutorial videos by Ben Orsburn link
  • Q Exactive Calibration $ Maintenance pdf