Vanquish Neo UHPLC System
The Vanquish Neo UHPLC system delivers maximum performance 24/7 for reproducible nano-, capillary-, and micro-flow LC-MS with a pressure rating of 1500 bar, here is a link to Thermo's website if you wish to learn more: Vanquish Neo UHPLC System
- Neo 1 (SN 8346660 and SN 8347102/ UW tag 2045623), Installed January/2023
- Neo 2 (SYS831576/ UW tag 2045624), Installed 8/7/2024
System specifications
Back to Top- System pressure
- up to 1500 bar (21,750 psi)
- Flow range
- 1 nL/min – 100 μL/min, in 1 nL increments
- 100 nL/min – 100 μL/min, recommended
- Injection volume range
- 0.01–25 μL
- Injection volume precision
- From 0.05 to 0.20 µL: ≤5.0% RSD
- From 0.20 to 0.50 µL: ≤1.5% RSD
- From 0.50 to 2.00 µL: ≤0.5% RSD
- Above 2 µL: ≤0.25% RSD
Plate configuration
Back to TopSample capacity (standard): 4 racks of 54 vials (≤ 1.5 mL) each
96-well plates with optional orientation barcoding to ensure the correct position in the autosampler and charger unit when loading and running the analytical sequence. Plates cannot be accepted in the wrong orientation.
Barcoded 96-well plate, V shaped well, 10-190 µl, 10pk 60180-217B (Fisher 03-255-015)
Barcoded 96-well plate, V shaped well, 10-190 µl, 100pk 60180-207B (Fisher 03-251-447)
Barcoded 96-well plate, U shaped well, 10-250 µl, 10pk 60180-216B (Fisher 03-255-014)
Barcoded 96-well plate, U shaped well, 10-250 µl, 100pk 60180-206B (Fisher 03-255-010)
You can also print/label the well plates yourself. Here are a couple of documents you can check out to start with:
- VanquishNeo_Barcode_96well_plates (docx)
- VanquishNeo_Barcode_A_96well_plates for ToughTag labels 1.28 x 0.5 in (e.g. TTLW-2016) (docx)
- VanquishNeo_Barcode_Z_96well_plates for ToughTag labels 1.28 x 0.5 in (e.g. TTLW-2016) (docx)
- VanquishNeo_Barcode_A_96well_plates for Avery labels 1.75 x 0.5 in (e.g. Waterproof Rectangle Labels with Sure Feed, .5" x 1.75", 80 White Labels per sheet (36545 or 36547)) (docx)
- VanquishNeo_Barcode_Z_96well_plates for Avery labels 1.75 x 0.5 in (e.g. Waterproof Rectangle Labels with Sure Feed, .5" x 1.75", 80 White Labels per sheet (36545 or 36547)) (docx)
Consumables
Back to Topthe following document includes part numbers for AS vials, columns trap cartdriges etc needed to use the VanquishNeo
HPLC setup
Back to Topthe following document includes guidelines to setup the capillary trap/column on our LC-MS systems, this is intended as a supplement to the hands-on training required to use the UWPR instruments.
Instrument Method setup guidelines
Back to TopDetailed guide how to setup the VanquishNeo method see pdf
Thermo Video: Simple and Intelligent Method Creation
Setup the Gradient Back to Top
- Note the buffer composition used on the VanquishNeo:
- Buffer A: 0.1 % Formic Acid in Water (Fisher Water with 0.1% Formic Acid (v/v), Optima™ LC/MS Grade, LS118)
- Buffer B: 0.1 % Formic Acid, 80% Acetonitrile in Water (Fisher Water with 0.1% Formic Acid (v/v), Optima™ LC/MS Grade, LS122500)
- Formula 100% ACN to 80% conversion: % B [80%ACN] = 1.25 * % B [100%ACN]
%B [100% ACN] | %B [80% ACN] | %B [100% ACN] | %B [80% ACN] | |
---|---|---|---|---|
1 | 1.25 | 10 | 12.5 | |
2 | 2.5 | 20 | 25 | |
3 | 3.75 | 30 | 37.5 | |
4 | 5 | 40 | 50 | |
5 | 6.25 | 50 | 62.5 | |
6 | 7.5 | 60 | 75 | |
7 | 8.75 | 70 | 87.5 | |
8 | 10 | 80 | 100 | |
9 | 11.25 | 90 | na | |
10 | 12.5 | 100 | na |
Installing SII for xcalibur and VanquishNeo software
Back to Top
Troubleshooting
Back to TopThis video from Thermo describes how to Replace the Sample Loop, Needle Unit, and Needle Seat Vanquish HPLC/UHPLC Maintenance Videos: How to Replace the Sample Loop, Needle Unit, and Needle Seat
This video from Thermo describes how to Replace the Pump Head Vanquish HPLC/UHPLC Maintenance Videos: How to Replace the Pump Head
This video from Thermo describes how to Replace the Piston Seals and Support Ring Vanquish HPLC/UHPLC Maintenance Videos: How to Replace the Piston Seals and Support Ring
This video from Thermo describes how to Replace the Check Valves Vanquish HPLC/UHPLC Maintenance Videos: How to Replace the Check Valves
TIPS AND TRICKS: ISSUE 1: How to avoid and recover from autosampler blockages (pdf)
Convert bar to psi
Back to TopFormula for bar to psi conversion: 1 bar = 14.5038 psi or (bar x 14.5038 = psi) or 0.0689475729 bars in 1 psi.
To convert bar to psi follow this link: UnitConverters.net: Convert bar to psi
For a bar / psi chart follow this link: VICI tech support: Pressure Conversion Chart
Convert Bar to | Psi | Convert Psi to | Bar | |
---|---|---|---|---|
1 | 14.5 | 10 | 0.7 | |
10 | 145.0 | 100 | 6.9 | |
50 | 725.2 | 500 | 34.5 | |
100 | 1450.4 | 1000 | 68.9 | |
150 | 2175.6 | 1500 | 103.4 | |
200 | 2900.8 | 2000 | 137.9 | |
250 | 3626.0 | 2500 | 172.4 | |
300 | 4351.1 | 3000 | 206.8 | |
350 | 5076.3 | 3500 | 241.3 | |
400 | 5801.5 | 4000 | 275.8 | |
450 | 6526.7 | 4500 | 310.3 | |
500 | 7251.9 | 5000 | 344.7 | |
550 | 7977.1 | 5500 | 379.2 | |
600 | 8702.3 | 6000 | 413.7 | |
650 | 9427.5 | 7000 | 482.6 | |
700 | 10152.7 | 8000 | 551.6 | |
750 | 10877.9 | 9000 | 620.5 | |
800 | 11603.0 | 10000 | 689.5 | |
850 | 12328.2 | 11000 | 758.4 | |
900 | 13053.4 | 12000 | 827.4 | |
950 | 13778.6 | 13000 | 896.3 | |
1000 | 14503.8 | 14000 | 965.3 | |
1050 | 15229.0 | 15000 | 1034.2 | |
1100 | 15954.2 | 16000 | 1103.2 | |
1150 | 16679.4 | 17000 | 1172.1 | |
1200 | 17404.6 | 18000 | 1241.1 |